Providing you the best range of linear shaker and bod incubator with effective & timely delivery.
₹ 56000 / Piece Get Latest Price
| Types | Horizontal shaker |
| Speed Range (RPM) | 40-200 |
| Country of Origin | Made in India |
| COA | Yes |
Minimum order quantity: 01 Piece
Description:
Linear Shaker is a laboratory device designed for uniform mixing, agitation, and shaking of samples in test tubes, flasks, or bottles. It operates with a back-and-forth linear motion, ideal for pharmaceutical, biotechnology, chemical, and microbiology laboratories. This device ensures consistent sample agitation for a wide range of applications, including dissolution studies, liquid mixing, and microbial growth.
| Parameter | Description |
|---|---|
| Motion Type | Linear Back-and-Forth |
| Speed Range | 50 – 300 RPM (Adjustable) |
| Platform Size | 300 x 300 mm / 400 x 400 mm (Custom Sizes Available) |
| Capacity | Up to 5 Kg or Custom Capacity |
| Speed Control | Digital / Analog |
| Timer | 0 – 99 Minutes or Continuous Mode |
| Display | LED Digital Display (Optional) |
| Power Supply | 220V, 50Hz |
| Body Material | Stainless Steel (SS 304) / Powder Coated Mild Steel |
| Motion Amplitude | 20 mm (Fixed / Adjustable) |
₹ 25000 / Piece Get Latest Price
| Number Of Shelves | 4 Shelves |
| Country of Origin | Made in India |
| Tempreature | up to 60 deg |
Minimum order quantity: 01 Piece
BOD incubator—standing for Biochemical (or Biological) Oxygen Demand incubator—is a specialized laboratory device designed for maintaining low and stable temperatures (typically between 5 °C and 60 °C) to support BOD testing, a key process in environmental water quality assessment. Unlike standard incubators that only heat, BOD incubators integrate both heating and cooling functions, making them essential in settings where precise, low-temperature control is indispensable.
Core Features & Design 1. Temperature Control & Uniformity· Range: Usually from around 5 °C up to 50–60 °C. Standard BOD Test Temp: Commonly maintained at 20 °C, though requirements may vary based on sample type (e.g., freshwater vs. wastewater microorganisms).
· Accuracy & Uniformity: Controllers (often PID with PT100 sensors) aim for ±0.2 °C to ±1 °C accuracy and uniformity.
2. Temperature Regulation Mechanisms· Heating: Via tubular or resistance wire heaters.
· Cooling: Uses compressors with CFC‑free refrigerants, or Peltier (thermoelectric) technology in some models.
3. Air Circulation· Uniform Heat Distribution: Achieved via axial fans or forced-air convection systems.
4. Construction Materials· Inner Chamber: Typically stainless steel (SS 304), resistant to corrosion and easy to clean.
· Outer Body: Powder-coated mild or cold-rolled steel; GMP models may feature full stainless construction.
· Insulation: High-density PUF or glass wool between walls for thermal efficiency.
5. Doors & Viewing· Many models feature an inner glass door (sometimes magnetic-sealed) for observing samples without disturbing the internal environment. 6. Controls & Safety
· PID Controllers: Provide accurate temperature settings; SV (set value) and PV (process value) are usually displayed. Some models include touchscreens, timers, or software integration. Safety Features: High/low alarms, over-temperature protection, and sometimes audible alerts.
7. Additional Functionalities· Illumination: Inside lighting (LED/fluorescent) often included, with optional timers.
· Interior Power Outlets: Provide power for stirrers, shakers, or BOD Trak systems.
· Mobility: Castor wheels allow easier repositioning.
· Data Logging & Compliance: Some advanced units support USB, RS‑232, PLC, or 21 CFR Part 11 compliance
Shanni Chaubey (Director)
Analysis Instruments
D-Mart, Shop No. 10, Sai Deep Plaza CHS, Yashwant Viva Township, Palghar
Vasai Virar - 401209, Palghar, Maharashtra, India